Device model | MDS-5V220A | Single channel Output power |
100W | communication method | TCP/IP (Network port) | |||
Discharge voltage | 5V-0.8V | Charging voltage | DC0-5V | data record | 100mS~65535 S | |||
Recharging current | 5Vseries:0.02A~20A | Discharge current | 5Vseries:0.02A~20A | |||||
Current accuracy | ±(0.05%RD+0.05%FS) | Voltage accuracy | ±(0.05%RD+0.05%FS) | |||||
Voltage resolution | When the lmV device detects or sets the voltage The smallest unit used | |||||||
Electric current resolution | 1mA device is detecting or setting voltage The smallest unit used | |||||||
Charging mode | Constant current charging Constant voltage charging Constant power charging | |||||||
Discharge mode | Constant current discharge Constant power discharge | |||||||
Charging cut-off condition |
Voltage,current,time, capacity,temperature | Discharge cut-off condition |
Voltage,current,time,capacity,temperature |
Input power | AC220±10%/50Hz (Cabinet power supply mode: three-phase five-wire) | ||
Main temperature | According to a particular apparatus 1 channel per channel optional | ||
input resistance | ≥10MQ | ||
Current response time | Current rise time ≤50mS | Power accuracy | ±(0.05%RD+0.05%FS) |
Number of steps in a single cycle | ≤100step,With nested loop function | ||
Circuit configuration | Modular structure design, easy replacement and maintenance | ||
VCELL,TCELL | Optional auxiliary temperature, auxiliary voltage, single module 8 channels, can be arbitrarily mapped to each main channel Temperature range: -40~140° (Voltage range: 0~5V/0~20V) |
performance | index parameter | |||
Input power | AC220V ±10%/50Hz | |||
Voltage |
Measuring range | 0.75~5V | ||
Output range | 10mV~5V | |||
Precision | ±0.1%*FSR | |||
Output accuracy | ±0.1%*FSR | |||
Current |
Measuring range | 0~20A | ||
Output range | 50mA~20A | |||
Measurement accuracy | ±0.1%*FSR | |||
Output accuracy | ±0.1%*FSR | |||
Power |
Measuring range | 0~900W | ||
Output range | 0~900W | |||
Measurement accuracy | ±0.2%*FSR | |||
Output accuracy | ±0.2%*FSR | |||
Data probability |
Sampling rate | 1HZ:1 samples/sec | ||
Recording rate | 1HZ:1 point/sec | |||
Recording conditions | T I V temp | |||
Full-scale current response time | ≤30ms | |||
Time resolution | 1s | |||
input resistance | ≥1MOhm | |||
Support charge and discharge type | Constant current charging, constant voltage charging, constant current and constant voltage charging, constant power charging, constant load discharging, constant power discharging, support channel paralleling | |||
Number of test commands | 399 commands can be set, 65535 loops can be set, 5 levels of nested loops are supported | |||
communication method | Ethernet,100Mbps | |||
safety protection | hardware | Add anti-reverse connection module to prevent battery reverse connection | ||
software | 1 Power-down data protection 2Support the setting of protection conditions: voltage upper limit, voltage lower limit, current upper limit, current lower limit, capacity upper limit, energy upper limit, voltage change rate, delay protection, temperature upper limit, temperature lower limit |
|||
Calibration frequency | ≥1year | |||
Fan noise | ≤85db | |||
Number of channels per chassis | 8aisles |